TriContinent C3000 / C24000

Category: Products > Pumps

    C3000 Pump

    C24000 Pump

    Drive Design

    Direct drive stepper motor/leadscrew with 30

    mm of travel. Manually moveable for power

    off syringe removal. Proprietary

    maintenance-free design.

    Direct drive stepper motor/leadscrew with 30

    mm of travel. Proprietary maintenance-free

    design.

    Stroke Speed

    1.2 seconds to 100 minutes per stroke

    (800 minutes in microstep mode).

    4.2 seconds to 800 minutes per stroke

    (3200 minutes in microstep mode).

    Step Resolution

    Standard mode 3,000 steps,

    Hi-Res mode 24,000 steps

    Standard mode 24,000 steps,

    Hi-Res mode 192,000 steps

    Part Number

    2703000

    2724000

    Power requirements:

    Supply voltage:

    24VDC 10%

    Currents:

    Max.1.5A

    Plunger drive:

    Principle:

    Quad encoder for loss of steps, detection of loss steps with error message

    Travel:

    30mm

    Plunger speed:

    1.2 sec. - 20 min./full stroke (speed ranges vary depending on syringe size and tubing)

    Syringes:

    Type:

    XP: 50µl, 100µl, 250µl, 500µl, 1.0ml, 2.5ml, 5.0ml

    Imprecision:

    0.1% CV at 50 L to 100 L

    0.05% CV at 250 µL to 12.5 mL

    Inaccuracy:

    1.0% at full stroke

    Fluid contact:

    Glass, PCTFE and Teflon

    Valve/valve drive:

    Valve options:

    120 3-port distribution

    Turn time:

    250ms between adjacent ports

    Drive:

    Stepper motor with optical sensor for position detection

    Positions:

    Input, output, bypass

    Materials:

    Plug: virgin teflon
    Body: PCTFE

    Fittings:

    1/4-28"-UNF screw fitting for syringe

    Interface:

    Type:

    RS 232, RS 485 or CAN bus

    Baudrate:

    9.600 or 38.400, 100K or 125K (CAN bus)

    Format:

    Data bits: 8 parity: none
    Stop bits: 1 half duplex

    Addressing:

    up to 15 XP/AL devices can be addressed individually

    Communication:

    Data terminal and OEM protocol

    Firmware:

    Programmable ramps, cutoffs, backlash compensation, plunger speeds, delays and loops
    Arbitrarily selectable plunger speed
    Clearance compensation
    Absolute and relative positions
    Diagnostics
    Flash-PROM

    Environmental:

    Temperature: 15-40C
    Humidity: 20-95%
    Non condensing RH at: 40C

    Dimensions:

    Height:

    127mm

    Width:

    44.4mm

    Depth:

    134.6mm